The device has been tested with a 100g sample cell.
The device has motor logic installed in to NICE and is driven by the new even-mode capable VIPER system
Principle:
Given that neutron instrument motors and detectors take several seconds to move and reach a stationary state to make the next data point measurable, we are developing a constantly rotating sample device that removes the deadtime associated with moving. For optimal samples, the continuous data collection and sample rotation can result in a double neutron count duty cycle.
We have recent experience adding a CHRNS: Cryogenic Goniometer to NICE and the motor control system, VIPER, was also updated to be compatible with time-stamps giving confidence in a swift incorporation of this device into the experiment control and data acquisition procedures.
System: Rotation Stage for 100mm ILL Cryostat
The rotation system consists of:
rotation stage to be mounted on top flange of the 100mm ILL cryostat
rotatable sample stick to be inserted in 100mm sample well
controller, cables
Specs for the rotation system:
Aperture diameter: 100+ mm
360 deg rotation with adjustable rotation limits
Bi-directional rotation
Position accuracy at sample position: 0.005 deg
Backlash at sample position: no more than 0.01 deg
Axial load capacity results from operation with vacuum in the sample well
Rotation speed MAX: no less than 2 deg/s
Rotation speed MIN: up to 0.1 arcmin/s
Distance from cryostat top flange to sample position: 1192mm
Nothing magnetic allowed within sample position, up to 1000mm above sample position
Sample stick material: 316 SS
Sample stick shall be ‘very stiff’ to avoid twist while rotation
The stick should be designed for continuous rotation back and forth with >1 year between service/failure
Calibrated, 1.4K– 300K, Cernox sensor attached at the sample stick bottom end
Sample mount: 5/16-18 threaded stud
Heat buffers to allow normal operation of the cryostat down to 1.5K at sample position
Sample stick shall be easily removable from the cryostat for sample change without removing the rotation stage mounted on top flange of the cryostat
Rotation system shall provide vacuum in the sample well of the cryostat, leak rate shall be better than 1e-8 mbar/lt sec