Vectorial second-harmonic magneto-optic kerr effect measurement

Published

Author(s)

Pavel Kabos, Anthony B. Kos, Thomas J. Silva

Abstract

A significant modification of an existing experimental technique based on the second-harmonic magneto-optical Kerr effect (SH-MOKE) is introduced. With a p-polarized pumping optical wave incident upon a magnetic film, the transverse component of magnetization causes a change in the second-harmonic generation efficiency of the material and the longitudinal component of the magnetization produces a change in the polarization rotation and/or ellipicity of the second-harmonic signal. This permits simultaneous hectorial measurement of the in-plane magnetization components. Examples of measured hysteresis loops from 50 nm thick permalloy films and procedures for SH-MOKE signal calibration are presented.
Citation
Journal of Applied Physics
Volume
87
Issue
9

Keywords

hysteresis, magnetic thin film, vectorial magnetizaition measurement

Citation

Kabos, P. , Kos, A. and Silva, T. (2000), Vectorial second-harmonic magneto-optic kerr effect measurement, Journal of Applied Physics, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=2481 (Accessed November 23, 2024)