We have developed a micromechanical torque sensor with sub-monolayer sensitivity for in situ monitoring of the magnetic moment of thin films during deposition. The film is deposited onto a microcantilever. The torque on the film is determined by measuring the deflection of the cantilever due to a small AC magnetic field perpendicular to the surface of the film. The microcantilevers have a high mechanical quality factor, large surface area, low spring constant, and high resonance frequency to improve film sensitivity to thickness. A phase-locked loop minimizes the resonance frequency shift of the cantilever due to mass loading and temperature drift that would otherwise affect the measurement of magnetic torque. The demonstrated sensitivity for a Ni0.8Fe0.2 film and for a Ni0.8Fe0.2/Cu multilayer film is less than 0.1 nm.
Citation
Journal of Magnetism and Magnetic Materials
Volume
286

Keywords

Cantilever, Fiber optic interferometer, Torque magnetometry, Ultra-high-vacuum

Citation

Min, D. , McCallum, A. , Russek, S. and Moreland, J. (2004), Micromechanical torque magnetometer with sub-monolayer sensitivity, Journal of Magnetism and Magnetic Materials, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=31680 (Accessed November 23, 2024)

Issues

If you have any questions about this publication or are having problems accessing it, please contact reflib@nist.gov.

Created October 28, 2004, Updated October 12, 2021