Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

Optimization of a Magnetostatic Cavity for a 3He Spin Analyzer on the CANDOR Polychromatic Reflectometer

Published

Author(s)

Md. T. Hassan, Wangchun Chen

Abstract

A New instrument, the CANDOR (Chromatic Analysis Neutron Diffractometer or Reflectometer) polychromatic beam reflectometer is being built at the National Institute of Standards and Technology Center for Neutron Research. CANDOR will have a capability of polarization analysis for both specular and off-specular scattering, and allow for measuring the magnetic structures of magnetic thin films with a significantly shortened measurement time as compared to a conventional monochromatic beam reflectometer at a continuous neutron source. A critical component of the polarization analysis capability is to develop a 3He neutron spin filter (NSF) for polarization analysis for both specular and off-specular scattering for a continuous wavelength range between 4 angstrom} and 6 angstrom} and a large solid angle coverage for up to 20 energy dispersive detector channels. For ex-situ spin-exchange optical pumping (SEOP) operation, a magnetically shielded solenoid (MSS) will be employed to provide a volume averaged field gradient, jr B?=Bj, better than 5 x 10-4 cm-1 to maintain ^3^He polarization relaxation times of at least a few hundred hours on the beam line even when there are significant stray fields from a magnet operating at a field up to a 3 T at the sample position. We describe a procedure to design, construct, and optimze a MSS with non-identical compensation coil dimensions on the ends to match the scattered neutron beam rays. The volume averaged field gradient is quantified and optimized over a cell for a MSS using the finite-element software package RADIA and compared with that from the experiment.
Proceedings Title
Journal of Physics: Conference Series
Volume
1316
Conference Dates
July 3-6, 2018
Conference Location
Milton Hill, GB
Conference Title
12th International Conference on Polarised Neutrons for Condensed Matter Investigations

Keywords

polychromatic reflectometer, CANDOR, SEOP, 3He neutron spin filter, field gradient

Citation

Hassan, M. and Chen, W. (2019), Optimization of a Magnetostatic Cavity for a <sup>3</sup>He Spin Analyzer on the CANDOR Polychromatic Reflectometer, Journal of Physics: Conference Series, Milton Hill, GB, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=927633 (Accessed December 26, 2024)

Issues

If you have any questions about this publication or are having problems accessing it, please contact reflib@nist.gov.

Created December 10, 2019, Updated October 12, 2021