Skip to main content
U.S. flag

An official website of the United States government

Official websites use .gov
A .gov website belongs to an official government organization in the United States.

Secure .gov websites use HTTPS
A lock ( ) or https:// means you’ve safely connected to the .gov website. Share sensitive information only on official, secure websites.

A Regional Comparison of Calibration Results for Type K Thermocouple Wire From 100 C to 1100 C

Published

Author(s)

K Garrity, Dean C. Ripple, M Araya, C R. Cabrera, L Cordova Murillo, M E. de Vanegas, D J. Gee, Edwin Guillen, S Martinez-Martinez, Edgar Mendez-Lango, L Mussio, S G. Petkovic, K N. Quelhas, G Rangugni, O Robatto, E Von Borrries Rocha

Abstract

Under the auspices of the Inter-american Metrology System (SIM), the National Institute of Standards and Technology (NIST) initiated a regional comparison for type K thermocouples from 100 deg. C to 1100 deg. C, with 11 participating countries. Each participating laboratory was sent new, unused wire from a lot of material characterized by NIST. The high uniformity eliminated any need to correct for variations of the transfer standard among the laboratories, greatly simplifying the analysis. The level of agreement among the laboratories results was quite good. Of the 380 total bilateral combinations of the data at the eight test temperatures, only 13 (i.e., 3.4 % of all combinations) are outside the k = 2 limits, and of these 13, only 3 are outside k = 3 limits. All of the outliers occur at temperatures of 800 deg. C and below, which suggests that drift of the type K wire due to high-temperature oxidation did not cause changes in thermocouple emf comparable to or larger than the claimed uncertainties.
Citation
TEMPMEKO
Volume
29
Issue
5

Keywords

comparison, temperature, thermocouple, type K

Citation

Garrity, K. , Ripple, D. , Araya, M. , Cabrera, C. , Cordova Murillo, L. , de Vanegas, M. , Gee, D. , Guillen, E. , Martinez-Martinez, S. , Mendez-Lango, E. , Mussio, L. , Petkovic, S. , Quelhas, K. , Rangugni, G. , Robatto, O. and Von Borrries Rocha, E. (2008), A Regional Comparison of Calibration Results for Type K Thermocouple Wire From 100 C to 1100 C, TEMPMEKO, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=831004 (Accessed November 21, 2024)

Issues

If you have any questions about this publication or are having problems accessing it, please contact reflib@nist.gov.

Created June 4, 2008, Updated October 12, 2021