The Industrial Thermometer Calibration Laboratory calibrates the following:
Calibrations in this laboratory are made by comparison with an ITS-90 calibrated standard platinum resistance thermometer in stirred liquid baths in the temperature range –196 ºC to 550 ºC.
Fixed-point measurements at the water triple point (0.01 °C) and the ice point (0 °C) are available.
How to Arrange a Calibration
Industrial Platinum Resistance Thermometers (IPRTs)
Temperature Range, °C |
U (k=2), mK |
---|---|
–196 |
2.3 |
–97 to –80 |
9.0 |
–80 to –70 |
4.0 |
–70 to 0.5 |
2.3 |
0.5 to 95 |
2.4 |
95 to 275 |
4.8 |
275 to 550 |
7.5 |
Thermistors
Temperature Range, °C |
U (k=2), mK |
---|---|
–196 |
1.8 |
–97 to –80 |
9.0 |
–80 to –70 |
3.6 |
–70 to 0.5 |
1.8 |
0.5 to 95 |
2.0 |
95 to 100 |
3.0 |
Thermocouples
Temperature Range, °C |
U (k=2), °C |
Thermocouple Type |
---|---|---|
–200 to 0 |
0.3 |
T |
–200 to 0 |
0.2 |
E |
–200 to 0 |
0.4 |
J |
–200 to 0 |
0.4 |
K & N |
0 to 500 |
0.5 |
K & N |
Digital Readout with attached probes †
Temperature Range, °C |
U (k=2), mK † |
---|---|
–196 |
2.3 |
–97 to –80 |
9.0 |
–80 to –70 |
4.0 |
–70 to 0.5 |
2.3 |
0.5 to 95 |
2.4 |
95 to 275 |
4.8 |
275 to 550 |
7.5 |
† The NIST calibration uncertainty does not include an uncertainty for the digital readout device.