Comparison of the Josephson Voltage Standards of the NIST and the BIPM

Published

Author(s)

Stephane Solve, Regis Chayramy, Michael Stock, Yi-hua D. Tang, June E. Sims

Abstract

A comparison of the 10 V Josephson array voltage standard of the Bureau International des Poids et Mesures (BIPM) was made with that of the National Institute of Standards and Technology (NIST), USA, in March 2009. The results are in very good agreement and the measured overall standard relative uncertainty is 1.2 parts in 10^10.
Citation
BIPM KCDB / Metrologia Technical Supplement

Keywords

CIPM-MRA key comparison, direct comparison, Josephson voltage standard, uncertainty

Citation

Solve, S. , Chayramy, R. , Stock, M. , Tang, Y. and Sims, J. (2009), Comparison of the Josephson Voltage Standards of the NIST and the BIPM, BIPM KCDB / Metrologia Technical Supplement, [online], https://tsapps.nist.gov/publication/get_pdf.cfm?pub_id=902739 (Accessed December 22, 2024)