Humidity Chamber

SANS humidity chamber

The SANS humidity chamber is a general purpose humidity generator that can control relative humidity from 5% to 95% Rh in the temperature range of 10 °C to 90 °C. The humidity is controlled by heating and cooling water wells within the chamber to generate or remove humidity from the cell. The sample temperature is controlled by circulating ethylene glycol and the wells are heated and cooled by peltier heaters controlled by a bipolar power supply. To promote condensation on the sample, the outer can is heated to above the sample temperature using band heaters.

Humidity Chamber Drawings

Specifications

Temperature Range 10 °C to 90 °C
Humidity Range 5% to 95% Rh
Can Dimensions 3.00" ID, 2.6" tall
Can Windows Aluminum, 2.00" tall, 0.020" thick

Contacts